Location -> Industrial online analyzer -> 2.Chlorine sensor

Type: Free chlorine
Illustrate:CXNS-238type residual chlorine is constant voltage sensor . It consist of two platinum electrodes and one micro-battery measurement system . When measuring ,the electrode measured end should keep in a stable electric potential .
CXNS-238 Free chlorine
(constant-voltage electrode)
·Application
Suitable for the residual chlorine in water industry and filing industry , chlorine dioxide or dissolved ozone measurements .
·Measuring principle
CXNS-238type residual chlorine is constant voltage sensor . It consist of two platinum electrodes and one micro-battery measurement system . When measuring ,the electrode measured end should keep in a stable electric potential . The different measured content will produce the different and good linearity current intensity under this electric potential .
·Feature:
★ Suitable zero capability , can make sure the accurately and reliably measurement .
★ Simple structure , easy to clean .
★ Suitable for the residual chlorine in the water ,chlorine dioxide ,ozone measurements .
·Technical parameter
Measurement range:0~2ppm、2~20ppm
Electrode:2 platinum rings
Reference electrode :GEL with Annular contacts
Material :glass
Electrode cable :3m high impedance and low noisy shielded cable
Operation pressure:10bar at 20℃(Max.)
Electrode size : Diameter 12mm ,length 110mm
(constant-voltage electrode)
·Application
Suitable for the residual chlorine in water industry and filing industry , chlorine dioxide or dissolved ozone measurements .
·Measuring principle
CXNS-238type residual chlorine is constant voltage sensor . It consist of two platinum electrodes and one micro-battery measurement system . When measuring ,the electrode measured end should keep in a stable electric potential . The different measured content will produce the different and good linearity current intensity under this electric potential .
·Feature:
★ Suitable zero capability , can make sure the accurately and reliably measurement .
★ Simple structure , easy to clean .
★ Suitable for the residual chlorine in the water ,chlorine dioxide ,ozone measurements .
·Technical parameter
Measurement range:0~2ppm、2~20ppm
Electrode:2 platinum rings
Reference electrode :GEL with Annular contacts
Material :glass
Electrode cable :3m high impedance and low noisy shielded cable
Operation pressure:10bar at 20℃(Max.)