Location -> Lab probes and sensors

Type:Laboratory Sodium Meter
Illustrate: CX-LSM-1609A Laboratory Sodium Meter CX-LSM-1609A is made based on the improvement of the previous pointer scale display, knobs, switches and non-intelligence design. It is equipped with the dedicated laboratory PPB sodium ion measurement elec..
CX-LSM-1609A Laboratory Sodium Meter
CX-LSM-1609A is made based on the improvement of the previous pointer scale display, knobs, switches and non-intelligence design. It is equipped with the dedicated laboratory PPB sodium ion measurement electrode and the reference electrode, with the static measurement performance in the beaker greatly improved, compared with the traditional sodium meters. As long as the user pays attention to the electrode cleaning, the satisfactory result can be achieved. It can be widely used for continuous monitoring of sodium ions of the solutions in thermal power plants, chemical fertilizer, metallurgy, environmental protection, pharmaceutical industry, biochemical industry, foodstuff and running water.
Features:
Double-row nixie tube: To display μg/L, pNa/mV and temperature simultaneously.
Computer-based: To adopt high-performance CPU chip, high-precision AD conversion technology and 8usMT
technology to complete multi-parameter measurement and temperature compensation, with high precision and
good repeatability
High reliability: Single board structure, touch button, no switch knob
Input impedance :> 1012Ω
High sensitivity: In the static water sample, it can detect 0.2ppb (8pNa)
Lasting stability: There is a very stable reading.
Waterproof and anti-erosion design
Network function: RS232 bi-directional communication, equipped with a printer or TLab8.0 communications and
data processing software.
Technical indexes:
Display: Double-row nixie tubes display
Measuring range: 0.2 ug/L~23000mg/L
0.00 Pna~8.00 PNa.
-1000~ +1000 mV. 0~99.9℃
Resolution: 0.1 ug/L; 0 .1 mV, 0.1℃
Repeatability: reading ± 2.5%
Accuracy: reading ± 2.5%, ±0 .2 mV, ±0.3℃
Continuous stability time: >100 seconds
Detection lower limit: 0 .2 ug/L (8pNa)
Input impedance: >1012Ω
Temperature compensation range: automatic/manual 0 ~ 99.9 ℃,
Response speed: 32 seconds (25℃) or 50 seconds (15℃)
Communication interface: RS232 bi-directional communication
Power supply: AC 220V ± 22V, 50Hz ± 1Hz, external power supply
Working conditions: ambient temperature: 0-60℃; relative humidity <≤90
Power-down memory: > 5 years
Calibration mode: One-point calibration, two-point calibration
Standard configuration:
One secondary instrument; one power supply; one reference electrode, one measuring electrode, one temperature
electrode; a set of electrode support, a set of standard buffer solution
Last:Laboratory Sodium Meter Next:Laboratory PH Meter