Location -> Products -> Analyzer Instrument -> Lab Instrument -> Lab Sodium Meter

Type:CX-LSM-1604 lab sodium meter
Illustrate:It can be widely used for the continuous monitoring of sodium ions in the solutions of the thermal power plants, chemical fertilizer, metallurgy, environmental protection, pharmaceutical, biochemical, foodstuff and running water.
CX-LSM-1604 Lab Sodium Meter
Brief Introduction
1.CX-LSM-1604 is a sodium meter with the most powerful functions and the most convenient operation on the market.
2.It is equipped with the dedicated laboratory ppb sodium ion measurement electrode and reference electrode.
3.The performance of the static measurement in the beaker has been greatly improved
Main Features
1.Computer-based It adopts CPU high-performance chip, high-precision AD conversion technology and SMT technology to complete multi-parameter measurement, temperature compensation and instrument self-checking
2.it is also featured by high precision and good repeatability.
3.High reliability Single board structure, touch button, no switch knob; input impedance> 1012Ω.
4.High Sensitivity In the static water sample, it can measure 0.2ppb (8pNa); With a long stablization time, it has very stable readings.
5.It is of waterproof and anti-erosion design.
6.Electrode diagnosis With a variety of calibration modes and EO and S queries, it conducts the monitoring of electrode performance.
7.Network function RS232 bi-directional communication, which can be equipped with a printer or TLab8.0communications and data processing software.
8.Historical data 3000 data can be stored.
9.Notepad 200 messages can be recorded.
10.Built-in real-time clock Providing a variety of functional time references.
Main Parameters
Measurement range | 0.2 ug/L~23000mg/L; 0~8.00 pNa, -1000~ +1000 mV 0~99.9°C |
Resolution | 0.1 ug/L; 0.1 mV,0.1°C |
Repeatability | Reading ± 2.5% |
Accuracy | ±0 2 mV, ±0 3°C |
Continuity stability time | >100 seconds |
Detection lower limit | 0.2 ug/L (8pNa); Input impedance >1012Ω |
Temperature compensation range | automatic/manual 0-99.9 °C |
Response speed | 32 seconds (25 °C) or 50 seconds (1 5°C) |
Communication interface | RS232 bi-directional communication |
Power supply | AC220V±22V, 50Hz ±1Hz |
Working conditions | Ambient temperature 0-60°C; relative humidity ≤90 |
power-down memory | > 5 years |
Clock accuracy | ± l minute/month |
Calibration mode | One-point calibration, two-point calibration, manual input of EO and s |
Last:Without the Next:CX-LSM-1609A laboratory sodium ..